The UTT810 relies on timing circuits whose response can change with temperature and operating conditions. Calibration support belongs to the resolved hardware contract. Startup/manual calibration and automatic temperature/time-trigger calibration are separate capabilities; a bootloader by itself does not establish either policy. Do not infer support merely from the presence of a C function or Python method.

Calibration calls are exposed in both the C API and the Python NexatomDevice class. The native library returns NEXATOM_ERROR_NOT_SUPPORTED (a NexatomError in Python) when the resolved image does not support an operation.

Manual calibration trigger

Calibration can be requested on demand through the native API. Perform it between acquisitions and inspect the device’s telemetry for its progress/completion. A successful trigger call means the command was accepted, not that calibration has finished; the SDK does not establish a universal completion time or a guarantee of uninterrupted measurement during calibration.

C API

nexatom_error_code_t nexatom_tt_trigger_calibration(nexatom_tt_handle device);

Python

device.enable_system(False)  # Stop the measurement before requesting calibration.
device.trigger_calibration()
# Observe the device's telemetry before starting the next measurement.

Auto-calibration configuration

Images with the established automatic-recalibration register contract can trigger it from elapsed time or temperature drift. The detailed settings below apply to those images. The resolved UTT_16_8_V1 application contract rejects configure_auto_calibration, enable_auto_calibration and set_calibration_settings with NEXATOM_ERROR_NOT_SUPPORTED (“Automatic calibration is not supported by the resolved modern device profile”) and issues no register writes. This restriction must not be generalized to every bootloader-based Zynq or Kintex image. A startup-calibration feature alone does not imply automatic recalibration support.

Trigger conditions

C API

Configure the thresholds:

nexatom_tt_configure_auto_calibration(
    device,
    5.0f,     /* 5.0 °C delta trigger */
    120       /* 120 minutes interval trigger */
);

Enable the subsystems:

nexatom_tt_enable_auto_calibration(
    device,
    true,     /* enable temperature trigger */
    false     /* disable time trigger */
);

Alternatively, apply all settings in a single call on a supporting image:

nexatom_tt_set_calibration_settings(
    device,
    5.0f,     /* temp_celsius */
    120,      /* time_minutes */
    true,     /* enable_temp */
    false     /* enable_time */
);

Python

# For an image supporting automatic recalibration; not part of normal startup.
device.set_calibration_settings(
    temp_celsius=5.0,
    time_minutes=120,
    enable_temp=True,
    enable_time=False,
)

The legacy temperature conversion uses the magnitude of a finite Celsius value, truncates to ADC codes and saturates at the 12-bit limit. The interval saturates at 4095 minutes; zero disables that threshold. Use reasonable experiment-specific values rather than relying on saturation. Configure thresholds before separately enabling triggers, or use the combined setter to apply them together.

Calibration status via telemetry

The legacy telemetry projection exposes calibration_metadata and temp_status_flags. Their bit definitions below are useful when inspecting a legacy image; modern applications should prefer the versioned telemetry view and its validity fields so an unavailable field is not confused with a false/zero state. See Telemetry and Diagnostics.

Calibration metadata (calibration_metadata)

The calibration_metadata byte contains 8 boolean flags mapped to the following constants:

Bitmask Constant Description
0x80u NEXATOM_TELM_CALIB_META_START_SEEN Calibration start sequence detected
0x40u NEXATOM_TELM_CALIB_META_DONE_SEEN Calibration completion detected
0x20u NEXATOM_TELM_CALIB_META_TEMP_TRIGGER_SEEN Last calibration was triggered by temperature
0x10u NEXATOM_TELM_CALIB_META_TIME_TRIGGER_SEEN Last calibration was triggered by timer
0x08u NEXATOM_TELM_CALIB_META_PENDING_LEVEL A calibration request is currently pending
0x04u NEXATOM_TELM_CALIB_META_ACTIVE_LEVEL The calibration engine is currently active
0x02u NEXATOM_TELM_CALIB_META_AUTO_ENABLED_LEVEL Auto-calibration is enabled in hardware
0x01u NEXATOM_TELM_CALIB_META_LAST_CALIB_AUTO The most recent calibration was automatic

Thermal warning flag (temp_status_flags)

The temp_status_flags byte in the telemetry struct includes a fixed thermal warning flag:

Bitmask Constant Description
0x02u NEXATOM_TELM_TEMP_STATUS_FIXED_WARNING Fixed thermal warning. Indicates the temperature has drifted by roughly ≥5°C from the baseline established at the last calibration.

On a legacy image, this warning can indicate that a new calibration should be considered between measurements. It is a firmware diagnostic threshold, not a thermal operating limit or a guarantee of restored precision. Modern images may report calibration/temperature differently through their versioned telemetry view.

Calibration data structure

The public SDK retains the following calibration record (NexatomCalibrationData in Python). The SDK still does not expose a public callback/getter that fills it after trigger_calibration(). An application must not allocate this record and treat its default values as a calibration result; use supported telemetry/status evidence instead.

nexatom_calibration_data_t

Field Type Description
calibration_time_ms uint64_t Milliseconds since the Unix epoch
is_valid bool True if calibration succeeded
validation_message char[256] Diagnostic message