The UTT810 relies on timing circuits whose response can change with temperature and operating conditions. Calibration support belongs to the resolved hardware contract. Startup/manual calibration and automatic temperature/time-trigger calibration are separate capabilities; a bootloader by itself does not establish either policy. Do not infer support merely from the presence of a C function or Python method.
Calibration calls are exposed in both the C API and the Python NexatomDevice class. The native library returns NEXATOM_ERROR_NOT_SUPPORTED (a NexatomError in Python) when the resolved image does not support an operation.
Calibration can be requested on demand through the native API. Perform it between acquisitions and inspect the device’s telemetry for its progress/completion. A successful trigger call means the command was accepted, not that calibration has finished; the SDK does not establish a universal completion time or a guarantee of uninterrupted measurement during calibration.
nexatom_error_code_t nexatom_tt_trigger_calibration(nexatom_tt_handle device);
device.enable_system(False) # Stop the measurement before requesting calibration.
device.trigger_calibration()
# Observe the device's telemetry before starting the next measurement.
0x0103); all inputs go dark together and come back together. On the UTT160810 the board holds its data (raw tags, count rates and Fast TIHI) for the whole calibration; its duration has not been measured and is estimated at about 1.2 s. Pulse-output settings survive a calibration, but triggers that arrive during it give no pulse.trigger_calibration() returns NEXATOM_ERROR_DEVICE_BUSY and writes nothing. A run, a live histogram or pulse outputs without a save do not refuse it.LIVE CALIBRATION_STARTED cause=manual src=sdk (log category LIVE_CHANGE).Images with the established automatic-recalibration register contract can trigger it from elapsed time or temperature drift. The detailed settings below apply to those images. The resolved UTT_16_8_V1 application contract rejects configure_auto_calibration, enable_auto_calibration and set_calibration_settings with NEXATOM_ERROR_NOT_SUPPORTED (“Automatic calibration is not supported by the resolved modern device profile”) and issues no register writes. This restriction must not be generalized to every bootloader-based Zynq or Kintex image. A startup-calibration feature alone does not imply automatic recalibration support.
temp_celsius).time_minutes).Configure the thresholds:
nexatom_tt_configure_auto_calibration(
device,
5.0f, /* 5.0 °C delta trigger */
120 /* 120 minutes interval trigger */
);
Enable the subsystems:
nexatom_tt_enable_auto_calibration(
device,
true, /* enable temperature trigger */
false /* disable time trigger */
);
Alternatively, apply all settings in a single call on a supporting image:
nexatom_tt_set_calibration_settings(
device,
5.0f, /* temp_celsius */
120, /* time_minutes */
true, /* enable_temp */
false /* enable_time */
);
# For an image supporting automatic recalibration; not part of normal startup.
device.set_calibration_settings(
temp_celsius=5.0,
time_minutes=120,
enable_temp=True,
enable_time=False,
)
The legacy temperature conversion uses the magnitude of a finite Celsius value, truncates to ADC codes and saturates at the 12-bit limit. The interval saturates at 4095 minutes; zero disables that threshold. Use reasonable experiment-specific values rather than relying on saturation. Configure thresholds before separately enabling triggers, or use the combined setter to apply them together.
The threshold the board uses is trunc(|ΔT| × 8.14) / 8.14 °C: 1 °C becomes 0.98 °C, 5 °C becomes 4.91 °C, 20 °C becomes 19.90 °C. A threshold of 0 °C fires on any temperature change.
Each call changes only its own part of the policy:
| Call | Changes | Keeps |
|---|---|---|
configure_auto_calibration() |
Temperature threshold and interval | The temperature and time enables in force |
enable_auto_calibration() |
The temperature and time enables | The threshold and interval in force |
set_calibration_settings() |
All four settings | — (it does not fire a calibration) |
Connect switches automatic calibration off on every image that has it, before control is granted, so a policy that an earlier session left on the board does not survive. Send your automatic-calibration policy after every connect, including after a re-plug on the same handle.
A calibration darkens the inputs, so automatic calibration waits while a file save of any kind runs and is caught up when the save ends:
NEXATOM_SUCCESS; the SDK stores the policy and writes it when the save ends.get_calibration_hold_state() (C: nexatom_tt_get_calibration_hold_state(), 64-byte record) reads the SDK’s own state and never touches the device:
| Field | Meaning |
|---|---|
held |
1 = the enables are held clear on the board |
due, due_reason |
1 = a calibration is due; NEXATOM_TT_CALIBRATION_DUE_TIME (1) or _TEMPERATURE (2) |
busy_reasons |
NEXATOM_TT_CALIBRATION_BUSY_RECORDING while a file save runs; “calibrate now” is refused exactly while this bit is set |
flags |
NEXATOM_TT_CALIBRATION_HOLD_FLAG_FAULT (the calibration stayed active too long), _BASELINE_VALID |
policy_word |
The policy last set in this session; 0 if none |
held_for_ms, since_last_start_ms |
How long the hold has lasted; time since the last calibration start (or connect) |
baseline_temperature_code, current_temperature_code |
Temperature codes at the last start and now |
incident_id |
The incident raised for a calibration fault; 0 without a fault |
The UTT160810 has no automatic calibration, so the call returns NEXATOM_ERROR_NOT_SUPPORTED there.
The legacy telemetry projection exposes calibration_metadata and temp_status_flags. Their bit definitions below are useful when inspecting a legacy image; modern applications should prefer the versioned telemetry view and its validity fields so an unavailable field is not confused with a false/zero state. See Telemetry and Diagnostics.
calibration_metadata)The calibration_metadata byte contains 8 boolean flags mapped to the following constants:
| Bitmask | Constant | Description |
|---|---|---|
0x80u |
NEXATOM_TELM_CALIB_META_START_SEEN |
Calibration start sequence detected |
0x40u |
NEXATOM_TELM_CALIB_META_DONE_SEEN |
Calibration completion detected |
0x20u |
NEXATOM_TELM_CALIB_META_TEMP_TRIGGER_SEEN |
Last calibration was triggered by temperature |
0x10u |
NEXATOM_TELM_CALIB_META_TIME_TRIGGER_SEEN |
Last calibration was triggered by timer |
0x08u |
NEXATOM_TELM_CALIB_META_PENDING_LEVEL |
Legacy telemetry layouts only: a calibration request is pending. Reserved and zero in current telemetry (TELM v10) |
0x04u |
NEXATOM_TELM_CALIB_META_ACTIVE_LEVEL |
The calibration engine is currently active |
0x02u |
NEXATOM_TELM_CALIB_META_AUTO_ENABLED_LEVEL |
Auto-calibration is enabled in hardware |
0x01u |
NEXATOM_TELM_CALIB_META_LAST_CALIB_AUTO |
The most recent calibration was automatic |
temp_status_flags)The temp_status_flags byte in the telemetry struct includes a fixed thermal warning flag:
| Bitmask | Constant | Description |
|---|---|---|
0x02u |
NEXATOM_TELM_TEMP_STATUS_FIXED_WARNING |
Fixed thermal warning. Indicates the temperature has drifted by roughly ≥5°C from the baseline established at the last calibration. |
On a legacy image, this warning can indicate that a new calibration should be considered between measurements. It is a firmware diagnostic threshold, not a thermal operating limit or a guarantee of restored precision. Modern images may report calibration/temperature differently through their versioned telemetry view.
The public SDK retains the following calibration record (NexatomCalibrationData in Python). The SDK still does not expose a public callback/getter that fills it after trigger_calibration(). An application must not allocate this record and treat its default values as a calibration result; use supported telemetry/status evidence instead.
nexatom_calibration_data_t| Field | Type | Description |
|---|---|---|
calibration_time_ms |
uint64_t |
Milliseconds since the Unix epoch |
is_valid |
bool |
True if calibration succeeded |
validation_message |
char[256] |
Diagnostic message |