The NexatomTT SDK includes a suite of Python example scripts located in the python/examples directory. These scripts are designed to provide reference implementations for common operational workflows, ranging from basic DLL initialization to advanced high-bandwidth data acquisition and firmware management.

The following tutorials provide step-by-step walkthroughs of these reference scripts, detailing the SDK classes used, the expected hardware state transitions, and the data processing pipelines.

Chapter contents

Topic Reference Scripts
DLL Load and Version Check load_smoke.py, version_info.py
Booting into Runtime from Bootloader boot_runtime.py
Realtime CPS and Telemetry Streaming ftdi_realtime_smoke.py
Raw Time-Tag Capture and Offline CSV Export file_save_and_offline_decode.py
Live TIHI and MFCO Plotting tihi_mfco_matplotlib.py
Runtime ↔ Bootloader Handoff Validation runtime_bootloader_handoff.py
End-to-End Firmware Field Update field_update_e2e.py

Processed and raw template anatomy

Start with python/examples/processed_acquisition.py for CPS/TIHI/MFCO CSV, or file_save_and_offline_decode.py for NXTT recording and CSV export. The quick start explains the settings and resulting files.

The C/C++ project in examples/sdk/ provides hardware.c and hardware.cpp, both sharing acquisition.c. C++ adds resource ownership around the public C ABI. The language-integration chapter gives build and run commands.

Stage What happens Where to adapt it
Enter runtime Discover the instrument by connection_id; native attaches or boots a valid runtime Runtime timeout and intended device selection
Validate Check the full plan against the native profile Selected channels, supported modes and features
Configure Quiet the system, configure inputs and measurement engines Threshold, edge, delay, hysteresis, TIHI span and coincidence window
Prepare recording Register callbacks and open native file sinks Destination, format and rotation limits
Acquire Start explicitly; callbacks retain owned snapshots while native records Observation duration and optional test pulses
Finish Stop, collect terminal results where appropriate, quiet sources, finalize files and check disconnect Preserve cleanup on every error path
Analyze Read new files and retain counts, units, completion and quality Small analysis functions and output summaries

The example sources include comments at these stages. A recorded histogram is not automatically an independent increment: the templates use REPLACE snapshots, so their summaries describe the latest result. Raw timestamps remain integer picoseconds through subtraction. These tutorials explain how those choices affect interpretation.